Aespire 7900
4-14
Electrical problems (power failure, etc.)
w
WARNING
If a circuit breaker opens frequently, do not use the system. Have a
qualified service representative repair the system
.
Symptom
Problem
Solution
Mains indicator is not ON.
The electrical power cable
is not connected.
Connect the power cable.
The inlet circuit breaker
(toggle switch) is open.
Close the circuit breaker
(Figure 4-1).
The power cable is
damaged.
Replace the power cable.
The electrical socket the
power cable connects to
has no power.
Use a different electrical
socket.
An internal fuse is open.
Have a qualified service
representative repair the
system.
One electrical outlet does not
have power.
The outlet circuit breaker is
open.
Close the circuit breaker.
A circuit breaker opens
frequently.
Equipment connected to
the outlet(s) (figure 4-1)
uses more current than the
circuit breaker rating.
Use a different power supply
for some of the equipment.
The equipment connected
to the outlet has a short.
Do not use the equipment
until it is repaired.
Tec 6 Plus vaporizer has no
power.
Not plugged into outlet.
Connect power cable.
The ventilator does not correctly
identify the breathing circuit
module.
Ventilator malfunction.
Ask a qualified service
representative to repair the
system.