1 Setup and Connections
1009-0633-000
1-7
w
WARNING
The filters must be in place to prevent dust and particles from
entering the breathing circuit.
6. When dry, place a new filter in the bottom of the canister, pour soda lime
into the canister and place a new filter over the soda lime before closing
and locking the cover. Wipe off any soda lime dust. Align the cover slots
with the canister locking tabs and press the cover down into place. Turn
the cover locking ring clockwise to lock the cover in place. Ensure cover is
properly sealed to prevent leaks and spillage. Alignment of the arrows
helps to indicate correct assembly.
7. When replacing the canister, make sure it is resting on both support pins
before latching it into place.
Pneumatic and electrical connections
w
WARNING
Equipment connected to the electrical outlets can increase the
leakage current. Regularly test the leakage current.
An optional isolation transformer kit is available to reduce total leakage
current.
w
CAUTION
Use only medical grade gas supplies. Other types of gas supplies
may contain water, oil, or other contaminants.
Mains inlet
Arrow shows the mains power inlet and cord.
Outlets
Labels show outlet voltage
ratings and circuit breaker amp ratings.