Aespire 7900
4-16
Pneumatic problems
w
CAUTION
No repair should ever be attempted by anyone not having
experience in the repair of devices of this nature.
Symptom
Problem
Solution
High-pressure leak test
fails.
Controls are not set correctly.
Set the system switch to
Standby and the auxiliary
flowmeter to OFF.
Incorrect cylinder connection (cylinder
yokes).
Make sure that there is only
one cylinder gasket, the
gasket is in good condition,
and the T- handle is tight.
Incorrect cylinder connection (DIN
connection.)
Make sure the nut is tight.
Low-pressure leak test
fails with a vaporizer on.
The vaporizer is not correctly installed.
Correctly install the
vaporizer.
The vaporizer filler is loose (fill port type
vaporizer).
Tighten the filler.
Vaporizer port o-rings (external) are damaged
or not installed.
Install new o-rings.
A vaporizer malfunction (the leak stops if you
use a different vaporizer in the same
position).
Send the vaporizer to a
Datex-Ohmeda Service
Center for repair.
A port valve malfunction (the leak continues
if you use a different vaporizer in the same
manifold position).
Have an qualified service
person repair the vaporizer
manifold.
Low-pressure leak with a
vaporizer OFF.
Anesthesia machine problem.
Contact a qualified service
representative.