Aespire 7900
4-6
Check Flow Sensors
Medium
(low after
acknowledged)
No flow or negative flow on
inspiratory sensor during
inspiration in a circle system
or negative flow on
expiratory sensor in
expiration (for 6 breaths in a
row).
Are the flow sensors correctly
installed?
Water build-up in the flow sensor
tubes?
Is a flow sensor tube cracked or
broken?
Inspect one way valves
(breathing circuit
module).
Replace flow sensor
module. Check the
condition of the flow
sensor and its tubing.
Circuit Leak Audio
Off
Low
Control setting on the Alarm
limit menu.
This message tells you that the
audio alarm for circuit leaks was
turned off.
- - -
Connect O
2
Sensor
Low
The O
2
sensor is not
connected to the cable.
Connect the sensor.
Contact a qualified
service representative
to replace the cable.
CPU Failure
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
CPU Internal Error
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
Display Voltage Out-
Of-Range
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
Exp Flow Sensor Fail
Low
The system cannot read the
calibration data stored in the
sensor.
Operation continues with default
values.
Replace the flow sensor.
- - -
Exp Reverse Flow
Medium
(low after
acknowledged)
Flow through the expiratory
sensor during inspiration (for
6 breaths in a row).
Look at the check valves.
Water build-up in the flow sensor
tubes?
Is a flow sensor tube cracked or
broken?
Replace the expiratory
check valve.
Check the condition of
the flow sensor.
Flow Valve (DAC)
Failure
Flow Valve (current)
Failure
Minimum
monitoring
(Medium)
Ventilator malfunction.
Ventilate manually. Monitoring is
still available.
Contact a qualified
service representative.
Gas Inlet Valve
Failure
Minimum
shutdown (High)
a
Ventilator malfunction.
Ventilate manually. Monitoring is
still available.
Contact a qualified
service representative.
Hardware Watchdog
Failure
Minimum
shutdown (High)
Ventilator malfunction.
Ventilate manually. Monitoring is
not reliable.
Contact a qualified
service representative.
Message
Priority
Cause
Action/Concerns
Repair