Aespire 7900
4-2
About alarms
w
WARNING
If an alarm occurs, safeguard the patient first, before
troubleshooting or repair procedures.
Two areas on the screen show alarms. The area at the top of the display shows
most alarms. If there are more than 4 alarms at the same time, the lower
priority alarms cycle every two seconds.
During severe malfunctions that prevent mechanical ventilation and/or
monitoring, the area under the waveform shows minimum system messages.
During normal operation, this area shows instructions (push the knob, etc.).
Alarm priority depends on the level of danger to the patient. High priority
alarms require immediate attention.
AB.90.025
Alarms
Priority
Alarm tone
Alarm silence
Note
High
10 tones,
10 second pause,
repeat
120 seconds or cannot
be silenced
Reverse video
Medium
3 tones,
25 second pause,
repeat
120 seconds
---
Low
Single tone
Tone does not repeat
---