Aespire 7900
3-10
1009-0633-000
The screen shows ‘Calibrating ... ‘, followed by the result: ‘Complete’ or ‘Failure’.
If the calibration fails,
•
do it again or
•
decrease the airway pressure. Then try again.
After repeated failures replace the O
2
sensor and re-calibrate at 21%.
Continue with steps 14 through 16 to exit to a normal screen.
Step 15
To exit, select “Go to Cal
Menu” and push the knob.
You can also press
the menu key to exit to
the normal screen
Step 16
Select “Go to Main Menu”
and push the knob.
Step 17
Select “Exit to Normal
Screen” and push the knob.
Step 18
Do a breathing circuit leak test before using the system.