2 Cleaning and Sterilization
1009-0633-000
2-5
•
One liter of this solution cleans four (4) flow sensors.
Procedure
1. Pull the latch to unlock the flow sensor module from the breathing
system.
2. Pull the flow sensor module from the breathing system.
3. Remove the flow sensors from the module.
•
Completely loosen the thumbscrew (1).
•
Pull off the flow sensor cover (3) from the flow sensor holder (2).
•
Remove the flow sensor connectors (4) from the flow sensor holder.
•
Pull the flow sensors (5) from the flow sensor holder.
1
2
3
4
5