3 User Maintenance
1009-0633-000
3-9
100% O
2
sensor
calibration
This procedure takes three minutes or less.
You must complete the 21% calibration before you can select the 100%
calibration.
w
WARNING
Do not perform calibration while unit is connected to a patient.
Step 11
Turn then push the knob to
select O
2
Sensor Cal.
Step 12
Select 100% then push the
knob.
Step 13
With the O
2
sensor in the
circuit, fill the circuit with
100% O
2
:
•
Push the flush
button.
•
Then
flow 100% O
2
at
5 L/min. (Circuit should
be open.)
Step 14
Select Start Cal. Then push
the knob to start calibration.