2 Cleaning and Sterilization
1009-0633-000
2-17
3. Remove the plugs from the ports. Permit the bellows to fully extend.
4. Close port 3.
5. Hold the bellows assembly upright. The bellows must not fall past the
guide line within one minute. If it does:
•
The port is not tightly sealed.
•
The bellows or the pressure relief valve is not correctly installed.
•
Parts are damaged.