Aespire 7900
6-6
1009-0633-000
Accuracy: At 20
°
C with gas supply pressures at 345 kPa (50 psi) and an
outlet pressure of 101.3 kPa (absolute) (14.7 psi) flowmeter accuracy agrees
with VDE 3513 Part 3, Accuracy Class 2.5 or better.
Different breathing circuit pressures, barometric pressures or temperatures
change the accuracy. With some conditions, these changes can be larger than
the tolerances.
Flush flow: 25-75 L/min.
O
2
supply failure alarm and shutoff:
IEC-60601-1 Clasification
The S/5 Aespire is classified as follows.
•
Class I Equipment
•
Type B Equipment
•
Ordinary Equipment
•
Not for use with flammable anesthetics
•
Continuous operation
•
Steam autoclavable or disinfectable per recommendations (see section
2
“Cleaning and Sterilization”
.
Electrical power
Supply voltage: 100-120 or 220-240 Vac
±
10% at 50 or 60 Hz.
Inlet Circuit Breakers:
Outlet Circuit Breakers:
System leakage current limit - do not exceed:
•
UL and CSA rated systems (USA and Canada): <300 µamps for the
system and all systems connected to electrical outlets.
•
IEC rated systems (Not USA and Canada): <500 µamps for the system
and all systems connected to electrical outlets.
Air
0.05 -0.95 L/min
1-15 L/min
O
O
O
O
2
2
2
2
PP
P
Prrrreeeessssssssuuu
urrrreeee::::
O
O
O
O
2
2
2
2
ssssuuu
uppp
pppp
pllllyyyy ffffaaa
aiiiilllluuu
urrrreeee aaa
allllaaa
arrrrm
m
m
m::::
193 to 221 kPa (28 to 32 psi)
N
N
N
N
2
2
2
2
O
O
O
O sssshhh
huuu
uttttooo
offffffff::::
3.5 kPa (0.5 psi)
Gas
Scale
(two flow tubes)
100-120 Vac
220-240 Vac
15A
8A
100-120 Vac
220-240 Vac
Japan
(3) 2A
(1) 3A
(3) 1A
(1) 2A
(2) 2A
(1) 4A