Aespire 7900
2-8
1009-0633-000
Disassemble the breathing system
The breathing system assembly can be disassembled for cleaning,
sterilization and part replacement.
1. Remove the breathing system and place on a flat surface.
2. Pull the latch to unlock the flow sensor module from the breathing
system.
3. Pull the flow sensor module from the breathing system.
4. Unscrew the sensor counterclockwise and remove it. To remove the
sensor cable from the breathing system, press on the connector button
(1) while pulling the connector out.
1