2 Cleaning and Sterilization
1009-0633-000
2-25
1. Pull the flexible gasket (1) from the receiver.
2. Pull the filter (1) out of its holder (2).
3. Do these steps in the opposite order to replace the filter and gasket after
autoclaving the receiver and gasket. Be sure the gasket is firmly pressed
into place at all points.
4. Before you use the system, complete the “
Preoperative Tests
” in section
4 of the User Reference Manual, Part 1.
Absorber canister
The absorber canister is available in two versions: Disposable Multi Absorber
and Reusable Multi Absorber. Both are removed and installed on the
breathing system in the same way.
Refer to
“Removing a canister”
in section 1
“Setup and Connections”
.
Only the Reusable Multi Absorber canister may be cleaned.
w
CAUTION
The filters must be in place to prevent dust and particles from
entering the breathing circuit.
1
1
2