4 Alarms and Troubleshooting
1009-0633-000
4-5
Aux Gas Outlet On
Medium
(low after
acknowledged)
The outlet selection switch is
set to the auxiliary common
gas outlet.
Connect the patient circuit to the
auxiliary outlet. For mechanical
ventilation or manual ventilation
with monitoring, select the
breathing system.
- - -
Backup Mode Active Low
SIMV-PC + PSV mode
entered.
Spontaneous breath rate fell
below the set breath rate
- - -
Battery Charger Fail
Low
The current in the battery
charging circuit is too high.
The system is operational, but
may fail later depending on what
caused this alarm.
Contact a qualified
service representative.
Battery Charging
Low
The battery is not fully
charged. If power fails, the
total backup time will be less
than 30 minutes.
Leave the system plugged in to
charge the battery.
- - -
Battery Current High
Low
Battery current >
6
amps
for
10 seconds.
The system continues to operate,
but may fail.
Contact a qualified
service representative.
Battery Failure High
Low
Battery voltage > 16 V for 10
seconds.
The system continues to operate,
but may fail.
Contact a qualified
service representative.
Battery Failure Low
Low
The battery voltage is too low
(<7 V) to supply the system if
power fails.
The battery does not have
enough charge to power the
equipment if power fails.
Leave the system plugged in to
charge the battery.
If the battery does not
charge in 24 hours,
contact a service
representative.
Cal Flow Sensors
Low
The last flow sensor
calibration failed.
Calibrate the flow sensors. Look
for water in the flow sensor
tubes. Dry if necessary.
Contact a qualified
service representative.
Calibrate O2 Sensor Low
O
2
%>110%
Does the sensor measure 21%
O
2
in room air?
Calibrate O
2
sensor.
Cannot Drive Bellows Low
The internal manifold
pressure is higher than Paw
+ tolerance.
Fill the bellows if empty.
- - -
Cardiac Bypass
Low
The alarm limit settings are
set for a patient on cardiac
bypass. Apnea alarms are
off.
Use the alarm limits menu to
change this setting.
- - -
Message
Priority
Cause
Action/Concerns
Repair