Aespire 7900
6-24
1009-0633-000
Fresh gas
compensation
Pressure
Volume
Flow compensation range:
200 mL/min to 15 L/min
Gas composition:
O
2
, N
2
O, Air, Anesthetic Agents
Patient airway pressure range:
-20 to +120 cm H
2
O
Patient airway display range:
-20 to +120 cm H
2
O
High pressure alarm set range:
12 to 100cm H
2
O, 1 cm increment
Sustained pressure alarm range:
6 to 30 cm H
2
O, 1 cm increment
Tidal volume display range:
5 to 9999 mL, 1 mL resolution
Setting range:
20 to 1500 mL
Minute volume:
0.0 to 99.9 liters
Breath rate:
4 to 100 bpm (breaths per minute), 1 bpm resolution
Volume sensor type:
Variable flow orifice