Aespire 7900
2-6
1009-0633-000
4. Submerge the flow sensor and
tubes in activated CIDEX
solution. Keep the connector dry.
5. Keep the solution in the tubes for
the sterilization period.
6. Submerge the flow sensor and
tubes in distilled water. Again, do
not get the connector wet.
7. Rinse as indicated in CIDEX
instructions.
8. Do steps 6 and 7 again to remove
all CIDEX.
9. COMPLETELY dry the flow
sensor and the tubes before
you use the sensor. Use a
dry syringe or connect
vacuum or pressure to
remove all liquid from the
sensor (sensor, tubes, and
connector):
w
CAUTION
Dry for > 1 min with these precautions:
• Maximum flow 10 L/min
• Maximum pressure ±100 cmH
2
O
10. Reverse steps 2 and 3 to reassemble the flow sensor module. Be sure to
align the flow sensor tubes with the grooves in the flow sensor holder.
11. Before you use the system, complete the “
Preoperative Tests
” in section
4 of the User Reference Manual, Part 1.
Remove the breathing system bag hose
1. Disconnect the bag hose (1)
from the bag hose connector
(2). Also remove the hose
from the clip (3).
2. If bag arm option is present,
remove the bag port elbow
from the bag arm support.
Push down on the release
latch and slide the bag port elbow out of the holder.
1
2
3