Aespire 7900
2-20
1009-0633-000
6. Insert the breathing circuit module into the bellows assembly aligned as
shown.
7. Rotate the breathing circuit module at the point shown by the dotted line
to attach it to the bellows assembly.
8. Make sure the o-ring (3) is on the O
2
sensor. Replace the sensor (1) by
screwing it in clockwise. Replace the O
2
sensor cable connector (2).
9. Attach the flow sensors to the module:
•
Insert the flow sensors (1) into the flow sensor holder. Note the
groove locations.
•
Attach the flow sensor connectors (2) to the flow sensor holder.
•
Attach the cover (4) to the flow sensor holder (3).
•
Tighten the thumbscrew (5) to fasten the cover.
2
1
3