Aespire 7900
3-2
1009-0633-000
Repair policy
Do not use malfunctioning equipment. Make all necessary repairs or have the
equipment serviced by an authorized Datex-Ohmeda service representative.
After repair, test the equipment to ensure that it is functioning properly, in
accordance with the manufacturer’s published specifications.
To ensure full reliability, have all repairs and service done by an authorized
Datex-Ohmeda service representative. If this cannot be done, replacement
and maintenance of those parts listed in this manual may be undertaken by a
competent, trained individual having experience in the repair of devices of this
nature.
w
CAUTION
No repair should ever be attempted by anyone not having
experience in the repair of devices of this nature.
Replace damaged parts with components manufactured or sold by Datex-
Ohmeda. Then test the unit to ascertain that it complies with the
manufacturer’s published specifications.
Contact your local Datex-Ohmeda Field Service Representative for service
assistance. In all cases, other than where Datex-Ohmeda’s warranty is
applicable, repairs will be made at Datex-Ohmeda’s current list price for the
replacement part(s) plus a reasonable labor charge.
Maintenance summary and schedule
These schedules are the minimum frequency based on typical usage of 2000
hours per year. You should service the equipment more frequently if you use it
more than the typical yearly usage.
User maintenance
Minimum
Frequency
Maintenance
Daily
•
Clean the external surfaces.
•
21% O
2
calibration (circuit O
2
sensor).
Weekly
•
Flow sensor zeroing.
Two weeks
•
Drain the vaporizers and discard the agent.
This is not necessary for Tec 6 vaporizers.
Summary of Contents for Aespire 7900
Page 10: ...Aespire 7900 viii 1009 0633 000 ...
Page 78: ...Aespire 7900 4 18 ...