6 Specifications and Theory of Operation
1009-0633-000
6-23
Volume control settings
•
V
T
(tidal volume)
•
Rate
•
I:E
•
P
limit
•
PEEP
Limited inspired pressure
mode
•
Selecting Pressure mode generates a pressure plateau linked to the
P
inspired
setting.
•
While observing the monitored P
max
value, the user adjusts P
inspired
until
P
max
is reached. The ventilator delivers a constant 30 L/m (inspiration)
during this mode.
•
P
insp
, Rate and I:E settings are used to create the breathing profile.
•
PEEP is available and is additive.
Pressure
mode settings
•
P
insp
(control pressure)
•
Rate
•
I:E
•
P
limit
(pressure limit)
•
PEEP
Minimum Monitoring
•
Shows data.
•
No mechanical ventilation.
•
“Monitoring Only” alarm message.
•
“Ventilate Manually” and specific failure message.
•
The software goes to minimum monitoring when a non-recoverable error
occurs in boot-up or normal operations.
Ventilation operating specifications
Pneumatics
Gas source:
Anesthesia system
Gas composition:
Medical Air or O
2
Nominal supply pressure:
350 kPa (51 psi)
Pressure range at inlet:
240 to 700 kPa (35 to 102 psi)
Flow valve range:
1 to 120 L/min at 240 kPa (35 psi)