2 Cleaning and Sterilization
1009-0633-000
2-3
Breathing system cleanable parts
1. Bellows assembly
2. APL Valve Ramp
3. APL Diaphragm
4. Breathing circuit module (O
2
sensor not autoclavable)
5. Absorber canister (reusable only)
6. Flow sensor cover
7. Flow Sensor Module (plastic flow sensors not autoclavable)
8. Exhalation Valve assembly
Figure 2-1 • Autoclavable assemblies
Parts marked
134°C
are autoclavable or washable by hand or machine (mild
detergent pH <10.5). Rinse and dry completely. All parts except the O
2
sensor
and disposable flow sensors can be washed.
If the flow sensors are plastic, refer to the
“How to clean and disinfect the flow
sensors”
procedure. If metal, they can be autoclaved at 134°C.
1
4
5
6
2
3
7
8