1 Setup and Connections
1009-0633-000
1-3
• Monitor the correct operation of all life support and monitoring
equipment.
• Keep backup manual ventilation available in case the
electrosurgical equipment prevents safe use of the ventilator.
• Do not use conductive masks or hoses.
Canister setup
w
w
w
w
WARNING
Obey applicable safety precautions:
w
Do not use the absorber with chloroform or trichloroethylene.
w
The Disposable Multi Absorber is a sealed unit which should not
be opened or refilled.
w
Avoid skin or eye contact with the contents of the absorber. In the
event of skin or eye contact, immediately rinse the affected area
with water and seek medical assistance.
w
Do not change the absorber during ventilation.
w
Inspect absorbent color at the end of a case. During non-use,
absorbent can go back to the original color. Refer to the absorbent
labeling for more information about color changes.
w
Desiccated (dehydrated) absorbent material may produce
dangerous reactions when exposed to inhalation anesthetics.
w
For systems with dual absorbent canisters, the carbon dioxide
absorbent material in both canisters shall be changed at least
weekly, preferably every Monday morning. For single canister
systems the absorbent material shall be changed every day,
preferably at the start of the day.