105
4)
The numeric readouts and bubble displays indicate the inclination of the leveling wafer.
•
The
Horizontal readings
(see figure below) can be displayed using different units and different conventions for
measuring the inclination. For more information, see “
Choosing Display Units and Conventions
Readings outside the working range of the leveling wafer are displayed as ####. Readings within the working
range, but beyond the range where readings meet the specified accuracy, are shown in gray. MultiView always
displays a fixed number of decimal places for readings, which might not reflect the actual accuracy of the
measurement. For information on the accuracy of the readings, see “
•
The
bubble level
displays (see the figure on the next page) show the inclination graphically in real time. In
both displays, the solid green circle represents inclination as a bubble in a bubble level would. The blue circle
indicates the Go/No-Go tolerance
, which you can adjust (see “
Setting the Go/No-Go Tolerance
In the left display, the perimeter represents the maximum inclination reading in the specified accuracy range
of the wafer, which depends on the wafer model (
see “
” on pag
). The display on the right is
a magnified view showing just the area within the dashed magnifier square of the display on the left. The size
of the magnifier square adjusts to keep the bubble in the right display visible.
Countdown for stabilizing
the readings