Tympanic Thermometer and Base
7
Table of Contents
Thermometer Display Icons and Alarms
The thermometer communicates to the user using the LCD display and audible beeps. When a probe cover is installed or the
batteries have been changed, the thermometer completes a system reset. The thermometer also performs a self-test to ensure
that the device is working correctly.
Alarm Condition
Display Mode
The LCD shows the low battery icon. The low battery icon will stay on until the
batteries are changed or until the dead battery display appears. Once the low
battery icon is shown, about 100 readings can be taken.
Dead Battery
The LCD shows the dead battery display. When any button is pressed this display
will flash 3 times and then the LCD will turn OFF. Once the dead battery icon is
shown, the batteries must be changed before the unit can be used.
Low Battery
Ambient Temperature below specified range
Ambient Temperature above specified range
Patient temperature below specified range
Patient temperature above specified range
If System Error “1” and System Error “2” are shown, allow the device to stabilize for 20 minutes before use. If System Error “12” is
shown, the site mode is corrupted. Refer to the image below for more information.
If System Error 2 is shown, the thermometer is out of calibration (e.g. a calibration
variable is outside of the expected range). Contact the service center.
If System Error “12” is shown, the site mode is corrupted. Allow the device to
power down and then enter Biotech Mode to reset the site mode. Save the
Biotech Settings before using the device again.
If System Error 1 is shown, the thermometer has an internal memory
checksum error (self-diagnostic test failure). Install a new probe cover to
reset the unit. If the system error does not clear, contact the service center.
For any other System Error, install a probe cover to reset the unit. If the system error does not clear, contact the service center.
Service addresses can be found in Section XI, Customer Service.
Summary of Contents for Genius 3
Page 206: ...Genius TM 3 Genius 3 21 4...
Page 209: ...el 2 Genius 3 5 VI XIII 14 17 PE...
Page 210: ...el 3 VI 10 X 30 IEC 60601 1...
Page 213: ...el 6 Genius 3 cardinalhealth com 1 2 VI 3 3 LCD LCD 4 5 6 7...
Page 214: ...el 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 215: ...el 8 LCD LCD 100 LCD 3 LCD 1 2 20 12 12 2 1 XI...
Page 216: ...el 9 C C F LCD C F 30 C F biotech 00 C C F F...
Page 267: ...Genius TM 3 Genius 3 21 4...
Page 270: ...ru 2 Genius 3 5 VI XIII 15 19...
Page 271: ...ru 3 VI 10 X 30 IEC 60601 1...
Page 274: ...ru 6 Genius 3 cardinalhealth com 1 2 VI 3 3 4 5 6 7...
Page 275: ...ru 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 276: ...ru 8 100 3 1 2 20 12 Site 12 Site Biotech Site Biotech 2 1 XI...