Tympanic Thermometer and Base
4
Table of Contents
Section IV — Instructions for Use
Peak Select System
The Genius 3 Tympanic Thermometer and Base uses the patented Peak Select System. It takes multiple readings and selects
the highest temperature for display.
Equivalence Modes
Genius 3 Tympanic Thermometer and Base is an ear canal thermometer for infants, children, and adults.
Prior to the introduction of tympanic thermometry, patient temperatures were measured in the mouth (oral), or in the rectum
(rectal). If a patient’s temperature was measured at the same time with each of these methods, different temperatures would
result. The Genius 3 Tympanic Thermometer and Base accounts for the average difference at each of these sites by adjusting the
displayed temperature.
The following equivalence modes are available on the Genius 3 Tympanic Thermometer and Base. Data is available from
Covidien on request.
Ear: In ear (EAR) mode, the display will indicate the absolute temperature without adjustment. The EAR mode is also the
UNADJUSTED MODE or DIRECT MODE of temperature acquisition.
Oral: In oral (ORL) mode, the ear temperature is adjusted to display the oral temperature equivalent.
Oral Mode = Ear Mode -0.09 °C
Clinical Bias = 0.09 °C
Limits of Agreement = (+/-) 0.64 °C
Clinical Repeatability = 0.13 °C
Reference Body Site = Oral Cavity
Measuring Body Site = EAR
Rectal: In rectal (REC) mode, the ear temperature is adjusted to display the rectal temperature equivalent.
Rectal Mode = Ear Mode +0.56 °C
Clinical Bias: = 0.5 °C
Limits of Agreement: = -0.47 / +1.66 °C
Clinical Repeatability: = 0.231 °C
Reference Body Site: = Rectum
Measuring Body Site: = EAR
Probe Covers
The Genius 3 Tympanic Thermometer and Base uses single use probe covers. The use of non-Covidien probe covers will cause
incorrect readings.
The probe covers are held in a cassette stored in the thermometer base. To load a probe cover on the thermometer, firmly push
the probe tip into the probe cover. When installed on the thermometer, the probe cover film should be smooth with no holes,
tears, or wrinkles. After the reading has been taken, eject the probe cover by pressing the eject button. Probe covers should
be disposed of properly after use. In order to help prevent and control infection, always use a new probe cover before taking
a reading.
Summary of Contents for Genius 3
Page 206: ...Genius TM 3 Genius 3 21 4...
Page 209: ...el 2 Genius 3 5 VI XIII 14 17 PE...
Page 210: ...el 3 VI 10 X 30 IEC 60601 1...
Page 213: ...el 6 Genius 3 cardinalhealth com 1 2 VI 3 3 LCD LCD 4 5 6 7...
Page 214: ...el 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 215: ...el 8 LCD LCD 100 LCD 3 LCD 1 2 20 12 12 2 1 XI...
Page 216: ...el 9 C C F LCD C F 30 C F biotech 00 C C F F...
Page 267: ...Genius TM 3 Genius 3 21 4...
Page 270: ...ru 2 Genius 3 5 VI XIII 15 19...
Page 271: ...ru 3 VI 10 X 30 IEC 60601 1...
Page 274: ...ru 6 Genius 3 cardinalhealth com 1 2 VI 3 3 4 5 6 7...
Page 275: ...ru 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 276: ...ru 8 100 3 1 2 20 12 Site 12 Site Biotech Site Biotech 2 1 XI...