Section VIII — Mounting Instructions
A wall mount or cart mount (sold separately) are available for the base unit. Follow the instructions supplied with those items.
Section IX — Troubleshooting
If the thermometer is not working properly, check the following:
Symptom
Action
Temperature reading unusually high
Check the probe cover for tears or gaps.
Temperature reading unusually low
Check the probe cover and thermometer tip for debris.
Check the patient ear canal for debris.
Low battery indicator lit
Replace the batteries.
Dead battery indicator lit
Replace the batteries.
Display blank
Replace the batteries.
System error displayed
If System Error “1” and System Error “2” is shown, allow the device to stabilize for
20 minutes before use. If System Error “12” is shown, the site mode is corrupted.
Allow the device to power down and then enter Biotech Mode to reset the site mode.
Save the Biotech Settings before using the device again. For all other system errors
the thermometer can be reset by installing a probe cover. If the system error does
not clear, send the thermometer for service. The service information is located in
Section XI, Customer Service.
The Genius 3 Tympanic Thermometer and Base advisories and alarms are described in the Thermometer Display Icons and Alarms
section of Section IV, Instructions for Use.
Section X — Specifications
Clinical accuracy characteristics and procedures are available from the manufacturer on request. To verify accuracy, use a certified
blackbody as specified in ISO 80601-2-56 or use a Genius Checker/Calibrator – order part number 303097.
Calibrated Accuracy Limits:
Ambient Temperature
Target Temperature
Accuracy
16°C to 33°C
33°C to 42°C
± 0.3°C
(60.8°F to 91.4°F)
(91.4°F to 107.6°F)
(± 0.5°F)
Calibrated Accuracy Limits (after Recalibration*):
Ambient Temperature
Target Temperature
Accuracy
16°C to 33°C
33°C to 42°C
± 0.3°C
(60.8°F to 91.4°F)
(91.4°F to 107.6°F)
(± 0.5°F)
*Post recalibration accuracy using the Genius Checker/Calibrator may not necessarily be equivalent to factory calibration.
Displayed Temperature Measurement Range:
Temperature Range depends on Site Mode as follows:
Mode
Range °C
Range °F
Ear
33.0 to 42.0
91.4 to 107.6
Oral
33.0 to 41.9
91.4 to 107.4
Rectal
33.6 to 42.0
92.4 to 107.6
Summary of Contents for Genius 3
Page 206: ...Genius TM 3 Genius 3 21 4...
Page 209: ...el 2 Genius 3 5 VI XIII 14 17 PE...
Page 210: ...el 3 VI 10 X 30 IEC 60601 1...
Page 213: ...el 6 Genius 3 cardinalhealth com 1 2 VI 3 3 LCD LCD 4 5 6 7...
Page 214: ...el 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 215: ...el 8 LCD LCD 100 LCD 3 LCD 1 2 20 12 12 2 1 XI...
Page 216: ...el 9 C C F LCD C F 30 C F biotech 00 C C F F...
Page 267: ...Genius TM 3 Genius 3 21 4...
Page 270: ...ru 2 Genius 3 5 VI XIII 15 19...
Page 271: ...ru 3 VI 10 X 30 IEC 60601 1...
Page 274: ...ru 6 Genius 3 cardinalhealth com 1 2 VI 3 3 4 5 6 7...
Page 275: ...ru 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 276: ...ru 8 100 3 1 2 20 12 Site 12 Site Biotech Site Biotech 2 1 XI...