![background image](http://html1.mh-extra.com/html/covidien/genius-3/genius-3_operating-manual_2673875011.webp)
en
Tympanic Thermometer and Base
9
Table of Contents
Site text
Pressing the °C/°F button when in this mode turns the body site text labels on or off. The labels will remain on when an “X” appears
inside the box icon, and the text will remain off when the box is empty.
Section V — Preventative Maintenance
A checker/calibrator is available for this device or the device can be sent in for service. This device must be checked for calibration
every 25 or 52 weeks, depending on your specific device. Refer to the display of your device. If three dots are visible above
the display, the device must be checked for calibration every 52 weeks. If no dots are visible, the device must be checked for
calibration every 25 weeks. For both devices, calibration is required whenever calibration is in question. If not able to calibrate,
contact your Covidien representative for details. Harsh use or harsh environmental conditions may result in the need for more
frequent checks. If the unit is dropped, abused, or stored at less than -25°C, or above 55°C, check it before the next use.
Refer to the display of your device. If three dots are visible above the display,
the device must be checked for calibration every 52 weeks.
Section VI — Cleaning & Disinfecting
Cleaning
Genius 3 Thermometer Body, Base, and Coil Cord:
A mild, common dish washing liquid detergent, such as Dawn®, should be used for general cleaning of the thermometer body,
base, and coil cord. This detergent should be used with a 20:1 ratio water to detergent mixture. The water and detergent mixture
should not exceed 55°C (130°F).
Caution:
The probe tip of the Genius 3 thermometer should not be cleaned with mild detergent.
Genius 3 Thermometer Probe Tip and Lens:
The probe tip and lens can be cleaned using a 70% Isopropyl alcohol wipe such as Webcol™*, Curity™*, or equivalent.
Caution: The use of other cleaners and disinfectants may cause significant damage to the Genius 3 thermometer and base
and may void warranty. Never use an abrasive pad on any surface of the Genius 3 thermometer.
Cleaning frequency:
It is recommended that the Genius 3 thermometer, base, and coil cord be cleaned after each use.
Summary of Contents for Genius 3
Page 206: ...Genius TM 3 Genius 3 21 4...
Page 209: ...el 2 Genius 3 5 VI XIII 14 17 PE...
Page 210: ...el 3 VI 10 X 30 IEC 60601 1...
Page 213: ...el 6 Genius 3 cardinalhealth com 1 2 VI 3 3 LCD LCD 4 5 6 7...
Page 214: ...el 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 215: ...el 8 LCD LCD 100 LCD 3 LCD 1 2 20 12 12 2 1 XI...
Page 216: ...el 9 C C F LCD C F 30 C F biotech 00 C C F F...
Page 267: ...Genius TM 3 Genius 3 21 4...
Page 270: ...ru 2 Genius 3 5 VI XIII 15 19...
Page 271: ...ru 3 VI 10 X 30 IEC 60601 1...
Page 274: ...ru 6 Genius 3 cardinalhealth com 1 2 VI 3 3 4 5 6 7...
Page 275: ...ru 7 8 9 10 10 C F C F C F 30 40 1 0 60 2 15 30 45 60 3 4 60 5...
Page 276: ...ru 8 100 3 1 2 20 12 Site 12 Site Biotech Site Biotech 2 1 XI...