102
SMARTFLOW II
Reloading Calibration Defaults (Cont.)
To confirm that defaults were actually loaded, the SmartFlow II will display LOAd for at least 1.5 seconds or as
long as the
CAL
or
DEC
buttons are held, up to 3 seconds. If the
CAL
and
DEC
buttons are held for 3 seconds, the
display will change to StorE and the defaults will be saved to EEPROM. If the
CAL
and
DEC
keys are held less than
3 seconds then StorE is never displayed and defaults are loaded but not stored in EEPROM.
NOTE:
When loading defaults the Tank value is loaded with the default ‘Tank Size’ to prevent
a Tank Alarm after loading defaults.
storE
LOAd
3 SEC
CAL
-
CALIBRATION