18
Intelect® Neo Clinical Therapy System
• Read, understand, and practice the precautionary and operating
instructions. Know the limitations and hazards associated with using
any electrical stimulation, laser device or ultrasound device. Observe
the precautionary and operational decals placed on the unit.
• All modalities should be routinely checked before each use to
determine that all controls function normally, especially that the
intensity control does properly adjust the intensity of the ultrasonic
power output in a stable manner. Also, determine that the treatment
time control does actually terminate ultrasonic power output when
the timer reaches zero.
• DO NOT use sharp objects such as a pencil point or ballpoint pen to
operate the buttons and touchscreen on the control panel.
• This unit should be operated at 10°C to 45°C and 0% to 90% Relative
Humidity. The unit should be transported and stored at 0°C to 60°C
and 0% to 95% Relative Humidity.
• Handle Ultrasound Applicator and Laser Applicator with care.
Inappropriate handling may adversely affect its characteristics.
• Before each use, inspect Ultrasound Applicator for cracks, which may
allow the ingress of conductive fluid.
• Inspect Applicator cables and associated connectors before each use.
• Device is designed to comply with electromagnetic safety standards.
This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with instructions,
may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a
particular installation. Harmful interference to other devices can be
determined by turning this equipment on and off. Try to correct the
interference using one or more of the following:
• Reorient or relocate the receiving device
• Increase the separation between the equipment
• Connect the equipment to an outlet on a different circuit from that
to which the other device(s) are connected and consult the factory
field service technician for help.
• Consult your authorized DJO dealer for help.
• Do not operate this unit when connected to any unit other than DJO
devices or accessories specifically described in user or service manuals.
• Use of controls, adjustments or performance of procedures other than
those specified herein may result in hazardous exposure to Laser
energy.
• DO NOT disassemble, modify, or remodel the unit or accessories. This
may cause unit damage, malfunction, electrical shock, fire, or personal
injury.
• Failure to use and maintain the Intelect® Neo System, its modules, and
its accessories in accordance with the instructions outlined in this
manual will invalidate the warranty.
• DO NOT permit foreign materials, liquids or cleaning agents to enter
the unit, including, but not limited to, inflammables, water, and
metallic objects from entering the unit, to prevent unit damage,
malfunction, electrical shock, fire, or personal injury.
• If you have difficulty operating the unit after carefully reviewing this
user manual, contact your DJO dealer for assistance.
• DO NOT remove the front and back covers. Doing so may cause unit
damage, malfunction, electrical shock, fire, or personal injury. There
are no user-serviceable parts inside the unit. If a malfunction occurs,
discontinue use immediately and consult dealer for repair service.
• Use of parts or materials other than DJO’s can degrade minimum
safety.
• The Intelect® Neo Clinical Therapy System is not designed to prevent
the ingress of water or liquids. Ingress of water or liquids could cause
malfunction of internal components of the system and therefore
create a risk of injury to the patient.
• The Intelect® Neo Vacuum Electrode Module is designed to operate
only when properly installed in the Intelect® Neo Clinical Therapy
System Cart.
• DO NOT operate the system in an environment of shortwave diathermy
use.
• Before each use, inspect Vacuum Electrode Cups and Lead Hoses for
cracks and damage which may not allow the vacuum to properly
secure the electrodes.
• Drain the Vacuum Electrode Module water reservoir regularly to
prevent excessive accumulation from electrode sponge water.
• Periodic flushing of the Vacuum system and the Vacuum reservoir are
required to maintain factory functionality of the Vacuum Electrode
Module. Refer to the Maintenance Section of this manual for proper
instructions.
!"#$%&'
!"#$%&'
PRECAUTIONARY INSTRUCTIONS