Intelect® Neo Clinical Therapy System
TENS- Symmetrical Biphasic
The Symmetrical Biphasic waveform has a short pulse duration and is capable
of strong stimulation of nerve
fi
bers in the skin and in muscle. This waveform
is often used in portable muscle stimulation units, and some TENS devices.
Because of its short pulse duration, the patient typically tolerates the current
well, even at relatively high intensities.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-73 mA (CC) 0-36 V (CV)
Phase Duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjustable 20-1,000 µsec
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-250 Hz
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Burst Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 bps
Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-250 Hz
Amplitude Modulation . . . . . . . . . . . . . . . . . . . . O
ff
, 40%, 60%, 80% and 100%
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
TENS- Alternating Rectangular
The Alternating Rectangular waveform is an interrupted biphasic current
with a rectangular pulse shape. This waveform is commonly used as a pain
management application.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-73 mA (CC) 0-36 V (CV)
Phase Duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjustable 20-1,000 µsec
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-250 Hz
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Burst Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-31 bps
Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-250 Hz
Amplitude Modulation . . . . . . . . . . . . . . . . . . . . O
ff
, 40%, 60%, 80% and 100%
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
WAVEFORMS
• Stimulus delivered by the TENS waveforms of this device, in
certain configurations, will deliver a charge of 25
microcoulombs (µC) or greater per pulse and may be
sufficient to cause electrocution. Electrical current of this
magnitude must not flow through the thorax because it
may cause a cardiac arrhythmia.
+"'*,)
IFC (Interferential) Traditional (4 Pole)
Interferential Current is a medium frequency waveform. Current is distributed
through two channels (four electrodes). The currents cross each other in the
body at the area requiring treatment. The two currents interfere with each
other at this crossing point, resulting in a modulation of the intensity (the
current intensity increases and decreases at a regular frequency).
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Carrier Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2000-10,000 Hz
Beat Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 sec
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep High Beat Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Scan Percentage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Static, 40%, 100% Manual
Amplitude. . . . . . . . . . . . . . . . . . . . 0-100 mA (CC with carrier freq
≤
5000 kHz)
. . . . . . . .. . . . . . . . . . . . . . 0-90 mA (CC with carrier freq > 5000 kHz)
. . . . . . . . . . . . . . . . . . . . . . 0-64 V (CV with carrier freq
≤
5000 kHz)
. . . . . . . . . . . . . . . . . . . . . . 0-45 V (CV with carrier freq > 5000 kHz)
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-60 Minutes
Available on Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1&2, 3&4 Option
TENS- Asymmetrical Biphasic
The Asymmetrical Biphasic waveform has a short pulse duration. It is capable
of strong stimulation of the nerve
fi
bers in the skin as well as of muscle tissue.
This waveform is often used in TENS devices. Because of its short pulse, the
patient typically tolerates the current well, even at relatively high intensities.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-93 mA (CC) 0-46V (CV)
Phase Duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adjustable 20-1,000 µsec
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-250 Hz
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Burst Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-10 Hz
Frequency Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-250 Hz
Amplitude Modulation . . . . . . . . . . . . . . . . . . . . O
ff
, 40%, 60%, 80% and 100%
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes