35
Intelect® Neo Clinical Therapy System
WAVEFORMS
Diadynamic Waveforms
The Diadynamic waveforms are recti
fi
ed alternating currents. The alternating
current is modi
fi
ed (recti
fi
ed) to allow the current to
fl
ow in one direction only.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1, 2, 3, 4
MF: (Monophasé Fixe) - Frequency of 50 Hz: phase duration of
10 ms followed by a pause of 10 ms.
DF: (Diphasé Fixe) - Frequency of 100 Hz: phase duration of 10 ms followed
immediately by another identical phase of 10 ms.
CP: (Modulé en Courtes Périodes) - 1 second of MF followed abruptly by 1
second of DF.
LP: (Modulé en Longues Périodes) - Rhythmical
fl
uctuation between 2 MF
currents.
CP-iso: (Courtes Periodes Isodynamic) - A combination of MF and DF
waveforms.
CP-id: Same as CP-iso.
MF+CP: A period of MF followed by a period of CP.
MF+CP-id: A period of MF followed by a period of CP-ID.
DF+LP: A period of DF followed by a period of LP.
DF+CP: A period of DF followed by a period of CP.
IFC Premodulated (Traditional 2 Pole)
Premodulated Current is a medium frequency waveform. Current comes out of
one channel (two electrodes). The current intensity is modulated: it increases
and decreases at a regular frequency (the Amplitude Modulation Frequency).
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . 0-100 mA (CC) 0-96 V (CV with carrier freq
≤
5000 kHz)
. . .. . . . . . . . . . . . . . . . 0-68 V (CV with carrier freq > 5000 kHz)
Carrier Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2,000-10,000 Hz
Beat Fixed (Sweep O
ff
) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep High Beat Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81-200 Hz
Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous or User De
fi
ned
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Min
Available on Channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1, 2, 3, or 4
Russian
Russian Current is a sinusoidal waveform, delivered in bursts or series of pulses.
This method was claimed by its author (Kots) to produce maximal muscle
strengthening e
ff
ects without signi
fi
cant discomfort to the patient.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA (CC) 0-90 V (CV)
Channel Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . .Single, Reciprocal, Co-Contract
Duty Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10%, 20%, 30%, 40%, 50%
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . O
ff
or On
Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous or User De
fi
ned
Burst Frequency (Anti-Fatigue O
ff
) . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-100 bps
Ramp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-5 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1, 2, 3, or 4
*CC= Constant Current
CV= Constant Voltage
VMS™ Burst
VMS Burst is a symmetrical biphasic waveform delivered in a burst format.
Because the pulse is relatively short, the waveform has a low skin load,
making it suitable for applications requiring high intensities, such as muscle
strengthening protocols.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-65 mA (CC) 0-32 V (CV)
Channel Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . .Single, Reciprocal, Co-Contract
Phase Duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-400 µsec
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . O
ff
or On
Set Intensity . . . . . . . . . . . . . . . . . . . . . . .Individual Channel Intensity Setting in
Reciprocal and Co-Contract modes
Cycle Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous or User De
fi
ned
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 bps
Ramp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-5 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1, 2, 3, or 4