26
Intelect® Neo Clinical Therapy System
SPECIFICATIONS
POWER (COMBINATION AND ELECTROTHERAPY
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 - 240 VAC, 2.5A to 1.25A, 50/60 Hz
Electrical Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CLASS I
Mode of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Electrical Type (Degree of Protection)
Ultrasound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
Laser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
Electrotherapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
Electrotherapy & sEMG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
Electrotherapy & Vacuum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
Ultrasound & Electrotherapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE B
NOTE:
All waveforms except High Voltage Pulsed Current (HVPC) have been
designed with a 200 mA current limit. VMS™, VMS™ Burst and all TENS
waveform output intensities are measured, speci
fi
ed, and listed to peak, not
peak to peak.
VACUUM SPECIFICATIONS
Vacuum Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 600 mbar maximum
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . (0-17.7 inches mercury maximum) +/- 5%
Vacuum Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Continuous or Pulsed
Continuous . . . . . . . . . . . . . 10 setting over vacuum range, 60mbar per setting,
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +10mbar to -10mbar per setting
Pulsed Mode
Maximum Vacuum settings 2 to 10, +10mbar to -10mbar per setting
Minimum Vacuum settings in 1 to 9, +10mbar to -10mbar per settingHold
Time in minimum & maximum vacuum settings,
0-20 seconds, in 1 second
steps, +/-0.5 seconds
Power
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-25 Vdc, maximum peak current 4A
Electrical Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CLASS I
Electrical Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TYPE BF
SYSTEM SPECIFICATIONS AND DIMENSIONS
Width
Depth
Height
Weight
Vacuum
10.39” (26.3906 cm)
5.11” (12.9794 cm)
11.07” (28.1178 cm)
4.9lb (2.2226 kg)
Module
11.12” (28.2448 cm)
6.34” (16.1036 cm)
1.43” (3.6322 cm)
1lb (0.453592 kg)
Head @ 45 degree with Base (Tabletop)
15.89” (40.3606 cm)
15.89” (40.3606 cm)
22.05” (56.007 cm)
20.7lb (9.389362 kg)
Cart Lowered (with casters)
23.94” (60.8076 cm)
26.19” (66.5226 cm)
27.41” (69.6214 cm)
29.4lb (13.33562 kg)
Cart Raised (with casters)
23.94” (60.8076 cm)
26.19” (66.5226 cm)
30.15” (76.581 cm)
Head and raised cart with screen @ 90deg
23.94” (60.8076 cm)
26.19” (66.5226 cm)
52.85” (134.239 cm)
48.9
lb (22.18067 kg)
GENERAL SYSTEM OPERATING AND STORAGE TEMPERATURE
Operating conditions
The device will meet its requirement under the following conditions:
Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10° C to 45° C
Relative Humidity: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0% to 90%
Atmospheric Pressure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 700hPa to 1060hPa
Transport and storage conditions
The device will remain in proper condition under the following conditions:
Temperature: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Above 0° C freezing to +60°C
Relative Humidity: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max 95%
Atmospheric Pressure: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 700hPa to 1060hPa