74
Intelect® Neo Clinical Therapy System
CLEANING THE INTELECT® NEO CLINICAL
THERAPY SYSTEM
With the system disconnected from the power source,
clean the system with a clean, lint-free cloth moistened
with water and mild antibacterial soap. If a more sterile
cleaning is needed, use a cloth moistened with an
antimicrobial cleaner.
Do not submerse the system in liquids. Should the unit
accidentally become submersed, contact the dealer or
DJO Service Department immediately. Do not attempt to
use a system that has been wet inside until inspected and
tested by a Trained Technician.
Do not allow liquids to enter the ventilation holes in the
optional modules. This could permanently damage the
modules.
Cleaning the LCD Screen
Clean the Therapy System LCD with a clean, dry cloth, in
the same way as cleaning the computer monitor screen.
Do not use abrasive materials or chemicals or liquids.
VACUUM MAINTENANCE
Cleaning the Vacuum Module
• With the system disconnected from the power source,
clean the vacuum module with a clean, lint free cloth
moistened with water and mild antibacterial soap. If a
more sterile cleaning is needed, use a cloth moistened
with an antimicrobial cleaner.
• Do not submerse the vacuum module in liquids.
Should the vacuum module accidentally become
submersed, contact the dealer or DJO Technical
Service Department immediately. Do not attempt to
use a vacuum module that has been wet inside until
inspected and tested by a Trained Technician.
Cleaning Instructions for the Electrodes and Suction
Cups:
• A mild antibacterial solution containing no chlorine
can be administered with a cloth and wiped or
air dried. This is recommended between patient
treatments. These electrodes are multiuse when
properly maintained and cleaned.
Cleaning Instructions for the Sponges:
• The associated sponges are recommended for single
patient use only and should be cleaned with a 70%
alcohol solution before and after each therapy session.
Reservoir Draining
• When draining the reservoir, wear surgical type gloves.
To drain the reservoir cup, rotate clockwise direction,
shown below. Dispose of contents in accordance with
your national, state or local disposal guidelines
Flushing Lead Hoses and reservoir
1. Connect all four lead hoses to the Vacuum Module.
Submerse the other end of the lead hoses into a
container filled with at least 250 ml (8 fl oz) of hot
water and one drop of dishwashing detergent added
2. Turn Vacuum Module ON and set Vacuum Intensity to
maximum
3. Repeat this procedure until no particles are visible
when the reservoir is drained
4. Dispose of reservoir contents according to national,
state and local rules and regulations
CALIBRATION REQUIREMENTS
Annual factory calibration is required for all Ultrasound
and Laser Applicators. Only the Applicators should be sent
to the factory or a Trained Technician for this procedure.
NOTE:
The unit was calibrated during the manufacturing
process and is ready to be placed into service upon
delivery.
Council Directive 2012/19/EU concerning Waste
Electrical and Electronic Equipment (WEEE)
requires not to dispose of WEEE as municipal
waste. Contact your local distributor for information
regarding disposal of the unit and accessories.
MAINTENANCE