79
Intelect® Neo Clinical Therapy System
ELECTROMAGNETIC COMPATIBILITY
(
EMC
)
TABLES
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Intelect Neo Clinical Therapy System is intended for use in the electromagnetic environment speci
fi
ed below. The
customer or the user of the Intelect Neo Clinical Therapy System should assure that it is used in such an environment
Emissions Tests
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The Intelect Neo Clinical Therapy System uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The Intelect Neo Clinical Therapy System is
suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fl
uctuations
IEC 61000-3-3
Compiles
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Intelect Neo Clinical Therapy System is intended for use in the electromagnetic environment speci
fi
ed below. The
customer or the user of the Intelect Neo Clinical Therapy System should assure that it is used in such an environment
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or ceramic tile.
If
fl
oors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2kV for power supply lines
±1kV for input/output lines
±2kV for power supply lines
±1kV for input/output lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV di
ff
erential mode
±2kV common mode
±1kV di
ff
erential mode
±2kV common mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0,5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for
5 sec
<5% U
T
(>95% dip in U
T
) for
0,5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for
5 sec
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the Intelect Neo Clinical Therapy System
requires continued operation during power
mains interruptions, it is recommended that
the Intelect Neo Clinical Therapy System be
powered from an uninterrupted power supply or
a battery.
Power frequency (50/60Hz)
magnetic
fi
eld
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fi
elds should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the a.c mains voltage prior to application of the test level.
APPENDIX