28
Intelect® Neo Clinical Therapy System
SPECIFICATIONS
LASER APPLICATOR TECHNICAL SPECIFICATIONS
For all single diode and cluster laser and LED applicators, the expected increase in the measured quantities after
manufacture added to the values measured at the time of manufacture is
±
20%.
The software incorporates a cooling function that forces the user to cool the laser cluster prior to the next treatment.
The software will calculate the cooling time needed when treatment times exceed 3 minutes per application. For a
3 minute treatment, it will force a 15 second cool down period before the next treatment can begin. For a 4 minute
treatment, it will force a 2 minute cool down period before the next treatment can begin. The software extrapolates for
times between 3 and 4 minutes.
A message will display on the screen informing the user that the probe is cooling down and the time period required.
After 5 seconds, this message will disappear. If the user attempts to use the probe before the cool down period is
completed, the message will re-display to signify that the applicator is still in cool down mode. After the cool down
period is complete, a message displays that informs the user that the unit is ready for use.
Single Diode Applicators
Applicator
Wave
length
(nm)
Output
Power
(mW0
Power Density
(W/cm
2
)
Treatment
Area(cm
2
)
Diode Type
Nominal Ocular
Hazard Distance
(NOHD-in meters)
Divergence
a1
(rad)
Divergence
a2
(rad)
Spot
Size
(cm2)
670 nm 10 mW LED
670
10
0.625
.016
LED
0.386308
0.698132
N/A
0.18
850 nm 40 mW Laser
850
40
0.571
0.07
Laser
2.4887
0.097
0.54334
0.012
850 nm 100 mW Laser
850
100
1.43
0.07
Laser
6.2218
0.097
0.54334
0.012
850 nm 150 mW Laser
850
150
2.14
0.07
Laser
8.80
0.097
0.54334
0.012
850 nm 200 mW Laser
850
200
2.85
0.07
Laser
12.44
0.097
0.54334
0.0376
850 nm 300 mW Laser
850
300
0.606
0.495
Laser
15.24
0.097
0.54334
0.242