41
Intelect® Neo Clinical Therapy System
DURA-STICK ELECTRODE INSTRUCTIONS
Connecting Lead Wires
1. Insert the lead with the Red (+) electrode connector
into one DURA-STICK Electrode
2. Insert the lead with the Black (-) electrode connector
into the other electrode
3. Make certain the lead wires are seated completely into
the electrodes
NOTE:
Use of conductive medium or sponges is not
required or recommended. DURA-STICK Electrodes are
manufactured to ensure the optimum conductivity
during therapy when properly applied.
Securing Electrodes
1. Remove the DURA-STICK Electrodes from the
protective backing
2. Apply to the treatment area as prescribed
3. Ensure the entire electrode surface is in contact with
patient skin by pressing into place
ELECTROTHERAPY PATIENT PREPARATION
ELECTRODE PLACEMENT
• Examine the skin for any wounds and clean the skin
• Apply the electrodes to the treatment area
• Ensure the electrodes are applied securely to the skin
• Ensure good contact between each electrode and the
skin
• Check the electrode contact regularly during the
treatment
• Examine the skin again after the treatment
• Choose electrodes that fit the anatomy
• View the Electrode Placement recommendations
in the Treatment Review screen for the particular
modality being used for treatment as a reference point
only prior to administering treatment
• Follow electrode manufacturer instructions
DURA
-
STICK™ ELECTRODES
DURA-STICK Electrodes are a self adhesive, disposable
product designed speci
fi
cally for use with Intelect® Neo
Clinical Therapy System.
It is recommended that DURA-STICK Electrodes be used
whenever possible to ensure the highest level of contact
with the treatment area and most uniform delivery of the
prescribed electrotherapy treatment.
For Electrotherapy operation, refer to page 50.
PATIENT PREPARATION