43
Intelect® Neo Clinical Therapy System
VACUUM ELECTRODE PATIENT PREPARATION
DJO Vacuum Electrodes are designed speci
fi
cally for use
with the Intelect® Neo Vacuum Module for optimum
e
ffi
ciency and performance of the Intelect® Neo Therapy
System. These electrodes are multi-use when properly
maintained and cleaned. The associated sponges are
recommended for single patient use only, should be
cleaned with a mild antibacterial solution containing no
chlorine, before and after each therapy session.
Vacuum Electrodes
Refer to the image below for available sizes of vacuum
electrodes. For waveform setup and parameter changes,
refer to page 52 Vacuum Operation. View the Electrode
Placement recommendations in the Treatment Review
Screen for the particular modality being used for prior to
administering treatment.
Connecting Lead Hoses
1. Insert the red lead hose WITH THE NARROW DIAMETER
HOUSING to the positive (+) lead hose connection
port for the channel desired on the Vacuum Module.
2. Connect the other lead hose, again with the NARROW
DIAMETER HOUSING to the negative (-) lead hose
connection port on the Vacuum Module.
NOTE:
Place the Vacuum plugs on all vacuum
channels not being used to prevent vacuum leakage
during a therapy session.
3. Select the prescribed Vacuum Electrode size and
install one to the opposite end of each Vacuum lead
hose.
The CORRECT connection, showing a SNUG fit of the
leads, WITH THE LARGER DIAMETER HOUSING, against
the electrode is shown here:
Compare the INCORRECT connection shown here:
.
4. Using cleaned sponges, liberally wet the respective
sponges, with distilled water, for the electrode size
being used. Squeeze out the excess water.
5. Place one sponge inside each Vacuum electrode cup.
6. Turn “ON” the Vacuum at the Treatment Review Screen.
30 mm
60 mm
90 mm
PATIENT PREPARATION