57
Function 0–Close All Valves and Turn All Pumps Off
This function closes all valves and turns all pumps OFF. This function also cancels Function 18. Always use Function 0 before exiting
the Diagnostics Function menu.
Note: To select multiple components, activate Function 18, then choose the
components.
Note: Before exiting the Diagnostics Menu, always choose Function 0 to deactivate all
devices, then choose Function 14 on Station B to open the water valve.
Function 1–Activate Detergent Valve
This function activates the detergent valve for the selected station.
Function 2– Activate Water Valve
This function activates the water valve on the main manifold for the selected station.
Function 3–Activate Alcohol Pump
This function activates the alcohol pump for the selected station.
Caution!
Do not activate the alcohol pump unless the alcohol valve is open. Perform
Functions 18, 12, 3 to activate.
Function 4–Activate Air System
This function activates the air system (compressor and air valve) for the selected station.
Caution!
Ensure the scope hookup is connected in the basin before performing this
procedure.
Function 5–Activate Air Valve
This function activates the air valve for the selected station.
Function 6–Activate Chamber Valve
This function activates the chamber valve for the selected station.
Function 7–Activate Disinfectant Pump
This function activates the disinfectant pump and valves 8 and 9 for the selected station.
Caution!
Ensure the scope hookup is connected in the basin before performing this
procedure.
Function 8–Activate Disinfectant Supply Valve
This function activates the disinfectant supply valve for the selected station.
!
!
!
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...