“Sheath Fail” is displayed (for optional leak tester only).
Large leak detected at beginning of cycle...........
Leak in scope. Allow the scope to reprocess, then remove scope and
send for repair. Leak between connectors. Press STOP to cancel the
cycle and open the lid locks. Check connection and reprocess scope.
Small leak detected during cycle ........................
Leak in scope. Allow the scope to reprocess, then remove and
manually leak test the scope.
Leak test adapter not connected..........................
Press STOP to cancel the cycle and open the lid locks. Connect correct
leak test adapter and reprocess scope.
Leak tester harness disconnected/defective ........
Check the leak tester harness connection. Replace a defective harness.
Leak tester board defective .................................
Replace the leak tester board.
Leak tester assembly defective ...........................
Replace the leak tester assembly.
“Shth Sen Err” is displayed during initial start (for optional leak tester only).
Scope is pressurized during startup ....................
Press STOP to cancel the cycle and open the lid locks.
Disconnect the leak tester hookup to release the pressure.
Reconnect the hookup.
Leak tester enabled, but option not present ........
Use Diagnostic 45 to disable leak tester option.
Leak tester board defective ................................. Replace the leak tester board.
Leak tester defective ........................................... Replace the leak tester.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...