INTERNAL 0.2 MICRON WATER FILTER–REMOVAL
The disinfector must be in idle state to perform this procedure.
1. Close the incoming water supply valve to the disinfector.
2. Drain excess water from the filter housing.
• Place a container under the water filter inlet tube.
• Disconnect the water filter inlet quick-connect from incoming water supply line.
• Connect the accessory hose to the water filter inlet tube.
• Open the filter bleeder valve and drain the water from the filter canister.
3. Remove the water filter cartridge.
Warning!
Always wear gloves when handling the filter.
• Loosen and remove the water filter housing.
• Rotate the filter counterclockwise to unlock.
• Remove the filter and discard in accordance with institution guidelines.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...