Wash Phase
The wash phase consists of up to two Soak segments. The operator has the choice to run one wash segment, two wash
segments or no wash segments. Setting the Soak time above zero using Setup 5, disables the second segment.
Note: If one or
two wash segments are chosen, then the flush phase is skipped.
During Soak, the detergent is injected into the basin through the scope channels according to the programmed detergent
injection time. The default detergent injection time is 3 seconds (~9ml). Please refer to the table in the Programming the
Reprocessor section for the appropriate dilution and timing settings. After injection, water fills the basin and the endoscope
is soaked in the basin for the programmed soak time. The basin is then drained while the endoscope channels are flushed
with fresh water. A Rinse Soak period then follows which is identical to the soak period. However, no detergent is injected
during this time.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...