![Cantel Medical MEDIVATORS DSD EDGE Service Manual Download Page 129](http://html1.mh-extra.com/html/cantel-medical/medivators-dsd-edge/medivators-dsd-edge_service-manual_3957444129.webp)
129
APPENDIX
Error Messages
Aborted .................................. Indicates a cycle was manually aborted and not complete
Air Disabled ............................ Indicates that the air flow sensor was disabled
A Dose Error ........................... Alarm indicating that the SSG Part A dosing reservoir has timed out and not all
...............................................
chemistry was delivered
Bas Disabled ........................... Indicates that the basin level sensor was disabled
Bas Sen Err ............................. Indicates that the basin level sensor is reporting a full basin at the start of a
...............................................
cycle (basin should be empty) or incoming water pressure is above 40 PSI
Basin Temp ............................. Indicates that the basin temperature has not reached the programmed
...............................................
minimum
B Dose Error ........................... Alarm indicting the SSG Part B dosing reservoir has timed out and
...............................................
not all chemistry was delivered
Clear Sys Alarm? ................... Diagnostic 70: prompts the user to press the enter button to clear a system
...............................................
alarm (such as a NVRAM error)
Flash High Error ..................... Indicates that the high code flash CRC test did not pass
Flash Low Error ...................... Indicates that the low code flash CRC test did not pass
Flo Disabled ............................ Indicates that the fluid flow sensor was disabled
Flow Sen Err ........................... Indicates that the flow sensor is reporting flow at the start of a cycle
...............................................
(no flow should be occurring)
Flow 2 Sen Err ........................ (SSG only) alarm indicating that the chamber fluid flow sensor is reporting flow
...............................................
at the start of the cycle (no flow should be occurring)
HLD Sampl Er ......................... Alarm indicating that a disinfectant sample port sensor is reading incorrectly
Lid ajar ................................... Indicates that the cover (lid) is open during a cycle
Lid Disabled ............................ Indicates that the cover (lid) opened sensor was disabled
Low Alcohol ............................ Indicates that the alcohol level is low
Low Chamber ......................... Indicates that the basin level did not reach the level sensor in time
Low Det. ................................ Indicates that the detergent level is low
No Air Flow ............................. Indicates that no air flow was detected
No Fluid Flo ............................ Indicates that no fluid flow was detected
No Fluid FI2 ............................ (SSG only) alarm indicating that no fluid flow was detected by the 3.2 l/min
...............................................
chamber fluid sensor during the mixing of the disinfectant.
NVRAM Err .............................. Non-Volatile RAM error, can only be cleared in Diagnostics 70
Power on ................................ Indicates that the device lost power during a cycle
RAM Error ............................... RAM error, can only be cleared in Diagnostics 70
Res T High .............................. Indicates that the incoming water temperature has exceeded the maximum
Res T Low ............................... Indicates that the temperature of the incoming water is below the minimum for
...............................................
safe disinfection
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...