Diagnostic Messages
Adjust Alarm ........................... Diagnostic 27: prompts the user to adjust the alarm volume.
Air Sen Enable ........................ Diagnostic 40: prompts the user to enable / disable the air flow sensor.
Alc. Sen Enable ....................... Diagnostic 47: prompts the user to enable / disabled the alcohol level sensor.
Auto Dis. Option ...................... Diagnostic 88: prompts the user to enable / disable the waterline disinfect option.
Bas. Temp Enable ................... Diagnostic 49: prompts the user to enable / disable the basin temperature sensor.
Basin Sen Enable .................... Diagnostic 42: prompts the user to enable / disable the basin level sensor.
Basin Temp. Cal ...................... Diagnostic 31: indicates that the DSD is in basin calibration mode.
Contrast adj ............................ Diagnostic 26: prompts the user to press the enter button to adjust the LCD contrast.
Count down from .................... Diagnostic 24: indicates that the DSD is performing a counter test.
CVR Sen Enable ...................... Diagnostic 41: prompts the user to enable / disable the cover (lid) open sensor.
Cycle Count ............................ Diagnostic 91: indicates that the DSD total cycle counts are being displayed.
Det. Sen Enable ...................... Diagnostic 48: prompts the user to enable / disable the detergent level sensor.
Diagnostics ............................. Indicates that the DSD is in diagnostic mode.
FLO Sen Enable ...................... Diagnostic 40: prompts the user to enable / disable the fluid flow sensor.
Heat Option............................. Diagnostic 88: prompts the user to enable / disable the heated reservoir option.
...............................................
Always set to 0 on the DSD EDGE
®
Endoscope Reprocessor
Input Code .............................. Prompts the user to enter the diagnostic entry password.
Input S/N major ...................... Diagnostic 82: prompts the user to enter the most significant two digits
...............................................
of the serial number.
Input S/N minor ...................... Diagnostic 82: prompts the user to enter the least significant two digits of the
...............................................
serial number.
Key Code ................................ Diagnostic 25: indicates that the following number is the ID number of the key just pressed.
Language ............................... Diagnostic 80: prompts the user to select a user language.
Max dis ................................... Diagnostic 71: Prompts the user to enter the maximum number of disinfectant
...............................................
cycles allowed.
Max. Bas. Temp ...................... Diagnostic 74: Prompts the user to enter the maximum water temperature allowed
Min. Bas. Temp ....................... Diagnostic 75: Prompts the user to enter the minimum temperature the basin has
...............................................
to be for disinfection.
Previous Key ........................... Diagnostic 25: indicates that the following number is the ID number of the key
...............................................
pressed before.
Prog. Flash? ............................ Diagnostic 85: prompts the user to press the enter button to start programming.
Programming .......................... Diagnostic 85: indicates that the program update is in progress.
Recirc Option .......................... Diagnostic 88: prompts the user to enable / disable the recirculation option.
Res High Temp En ................... Diagnostic 49: prompts the user to enable / disable the water high temperature monitor.
Res Low Enable ...................... Diagnostic 43: prompts the user to enable / disable the low reservoir sensor.
...............................................
Must be set to 0 on the DSD EDGE Endoscope Reprocessor
Res Low Temp En ................... Diagnostic 49: prompts the user to enable / disable the water low temperature monitor.
Res. Temp. Cal. ....................... Diagnostic 32: indicates that the DSD is in water calibration mode.
Reset programs? ................... Diagnostic 72: prompts the user to press the enter button to clear all user programs.
Set A purge Secs . .................. Diagnostic 63: prompts the user to enter the air purge time in seconds.
Set D Drain Secs ..................... Diagnostic 61: prompts the user to enter the disinfectant drain time in seconds.
Set D Fill Secs ........................ Diagnostic 62: prompts the user to enter the disinfectant fill time in seconds.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...