88
Air Phase
The Air phase is simply a programmed time during which air is purged through the endoscope channels.
Note: Please refer to the Disinfection Cycle Chart for user programmable time settings, default
state times, and time limitations.
Note: Every drain transaction is followed by an air purge where air is flushed through the
endoscope channels and internal fluid lines. The Air LED on the control panel blinks during air
purge.
Note: During each phase of the cycle, an LED illuminates to indicate the present phase. The Flush
LED indicates that one of the wash phase segments is running.
Pre-start Inspection
Use the following procedure to inspect the reprocessor before startup.
1 Check the external pre-filters if installed on the incoming water supply. Replace any filters if the dynamic
pressure drop across any filter is greater than 10 psi. Inspect the pressure gauges for pressure reading when water
is flowing through the filters.
2 Check the detergent reservoir (if utilized) for proper detergent level. Add detergent, if necessary.
3 Check the alcohol reservoir (if utilized) for proper alcohol level. Add alcohol, if necessary.
CAUTION!
For the alcohol and detergent bottles; remove, refill and replace each bottle
(alcohol and detergent) separately to ensure the correct fluid is in reservoir.
!
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...