63
Function 25–Test Keypad
Use this function to perform a keypad test.
• Press each key to verify the correct key code displays.
• Press and hold the “*”, then press the “#” key to stop the test
Key
Code
Key
Code
1
01
ID Data
18
2
02
Heater On
15
3
03
Enter
10
4
04
Program
19
5
05
Disinfectant Dump
16
6
06
Cancel
11
7
07
Add Air
20
8
08
Set Up
17
9
09
Station Select
14
0
00
Start
12
*
21
Stop
13
#
22