Process Interruption
A process interruption may occur due to a system interruption, or initiated by the operator.
System Interruption
A system interruption may be caused by loss of water or air, loss of power, or by a lack of available Part A and Part B
disinfectant components.
Note: During an operating cycle, reprocessors with optional lid locks cannot be opened
except by means of a service code.
1. Correct the error, then press the START button.
- The reprocessor continues the cycle from the point of interruption.
- If the interruption is caused by a power outage, the reprocessor automatically restarts the cycle when power
is restored. A “Power On” message will be indicated in the log.
Operator Initiated Interruption
1. Press the STATION SELECT button to choose station A or station B.
2. Terminate Cycle: Press the CANCEL button, then the ENTER button. The present cycle is aborted. The reprocessor
fails to a safe mode and the endoscope must be reprocessed.
- If the cycle is terminated, an “Aborted” message will be displayed. The endoscope should not be used unless
a “Cycle Completed” message is displayed.
Service: If the error recurs, or cannot be corrected, refer to the Troubleshooting chapter.
3. Interrupt Cycle: Press the STATION SELECT button, then the STOP button.
- To resume the cycle press the START button. The cycle proceeds as normal.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...