“No Fluid Flow” is displayed during disinfect phase.
Disinfectant screen is clogged............................
Clean the disinfectant screen.
Disinfectant supply valve defective....................
Repair or replace the component.
Scope to basin connection disconnected............. Reconnect and restart the cycle.
Hookup is pinched or kinked.............................
Check for pinched or kinked hookup. Cancel the cycle, then reposition
the hookup. Restart the cycle.
Pump does not operate........................................ Check that the disinfectant filter is connected and that pump is primed.
Replace if necessary.
Scope channel is blocked.................................... Remove scope and send for repair.
Sensor harness disconnected/defective............... Check the sensor harness connection. Replace a defective harness.
Sensor board defective........................................ Replace the sensor board.
Sensor defective.................................................. Verify proper fluid flow. Replace the sensor.
“No Fluid Flow” is displayed during flush or rinse phase.
Scope to basin connection disconnected............. Reconnect and restart the cycle.
Water pressure too low........................................ Pressure must be between 35-40psi (2.4 - 2.75bar).
Prefilter/internal filter clogged............................
Change the filter.
Air lock in the water filter...................................
Open the bleeder valve on the filter to purge air.
Water filter is clogged.........................................
Check the water pressure at inlet and outlet regulators. Lower than
normal pressure may indicate a clogged filter. Replace the filter.
“No Fluid Flow” is displayed during flush or rinse phase.
(continued)
Air pressure switch harness disconnected........... Check the harness for proper connection. Also, check for damage,
loose wires.
Sensor harness disconnected/defective............... Check the sensor harness connection. Replace a defective harness.
Sensor board defective........................................ Replace the sensor board.
Sensor defective.................................................. Verify proper fluid flow. Replace the sensor.
“No Fluid FI2” is displayed.
Disinfection filter clogged ................................... Clean filter.
Restriction in drain.............................................. Clean drain screen in basin.
“Res T High” is displayed.
Incoming water pressure to high ........................ Adjust incoming water temperature.
Incorrect program temperature setting................ Use Diagnostics 74 to enter value.
Thermistor defective ........................................... Repair or replace the component.
“Res T Low” is displayed.
Incoming water pressure to low ........................
Adjust incoming water temperature.
Incorrect program temperature setting................ Use Diagnostics 74 to enter value.
Thermistor defective ........................................... Repair or replace the component.
Heater turned off or defective ............................. Turn heater on or replace. Check harness connection.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...