116
FLOW SENSOR–CLEAN
1. Locate the flow sensors.
2. Remove the flow sensor and disassemble.
• Disconnect the sensor cable from the sensor board.
• Release the CPC fitting at the bottom of the flow sensor.
• Remove the flow sensor assembly by releasing the tube coupling.
• Remove the plug from the flow sensor assembly.
• Remove the plunger from inside the flow sensor assembly.
3. Clean the flow sensor.
• Wipe the plunger using a clean cloth.
• Push a clean cloth through the barrel of the flow sensor.
4. Re-assemble the flow sensor.
• Replace the plunger into the flow sensor with the flat end in first.
• Inspect the O-rings of the plug, side coupling and bottom coupling for damage.
• Apply silicon grease to each O-ring and reassemble.
5. Reinstall flow sensor.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...