125
No air after cycle.
Incorrect air program setting...............................
Check the air setting in the Setup menu (Setup 5).
Excessive fluid remaining in endoscope channels
Check for correct compressor operation.
LCD screen unreadable.
Contrast out of adjustment..................................
Use the Diagnostic function 26 to adjust contrast.
Ribbon cable disconnected, damaged.................
Check the ribbon cable between the LCD screen and the control panel.
Reconnect a loose cable, or replace a damaged cable.
LCD screen failure..............................................
Replace the control panel.
Printer does not operate.
No power to reprocessor .....................................
See “Reprocessor does not start”.
Power supply unplugged.....................................
Check the power supply connection is plugged into the outlet in
the reprocessor.
Printer ribbon is jammed.....................................
Check the printer ribbon. Open the printer and re-align the jammed
ribbon. Rotate the knob to adjust the ribbon tension.
Printer paper is jammed ......................................
Check the printer paper. Open the printer and
remove the paper jam. Press the paper feed switch
to feed paper through the printer cover.
Printer connection is disconnected .....................
Check the printer cable connection. Verify the connector is
plugged in and seated tightly. Replace a damaged cable.
Printer damaged or defective ..............................
Replace the printer.
Defective motherboard .......................................
Replace the motherboard.
Printing unreadable.
Printer ribbon is infrequently used......................
Press the paper feed switch to advance the printer ribbon
to a new section.
Printer ribbon worn.............................................
Replace the printer ribbon.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...