Function 9–Activate Disinfectant Overflow Valve
This function activates the disinfectant overflow valve for the selected station.
Function 10–Activate Sample Port Valve
This function activates the disinfectant sampling valve.
Function 11–Activate Drain Valve
This function activates the drain valve for the selected station.
Note: To transfer fluid from the basin to the waste drain, activate Function 11.
Function 12–Activate Alcohol Valve
This function activates the alcohol valve for the selected station.
Function 13–Activate Isolation Valve
This function activates the Isolation Valve.
Function 14–Activate Water Inlet Valve
This function activates the water inlet valve for the entire reprocessor. This function is activated from station B
diagnostics only.
Note: Before exiting the Diagnostics Menu, always choose Function 0 to deactivate all
devices, then choose Function 14 on Station B to open the water valve.
Function 15–Activate Disinfectant Inlet Valve
This function activates the disinfectant inlet valve. The disinfectant inlet valve is used to supply the 0.2 micron filter during
automatic water line disinfection procedures.
Function 16–Activate Recirculation Inlet Valve
This function is not available on the DSD EDGE
®
Endoscope Reprocessor.
Function 17–Activate Recirculation Chamber Valve
This function is not available on the DSD EDGE Endoscope Reprocessor.
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...