DISPLAY STATE TIME
A cycle is comprised of a number of states. Use the following procedure to view the state time for both stations.
1. Press the SETUP button.
• Enter 18 on the keypad, then press the ENTER button.
2. The current state number and time remaining for each station is displayed
3. Press the SETUP button to exit the display.
Note: Refer to the Disinfection Cycle Chart in the Appendix for state times.
Figure 18: State Time Screen
Station B
Station A