90
Disinfecting Endoscopes
Use the following procedure to prepare an endoscope for disinfecting, to run the disinfection process, and to complete the
disinfection process.
Preparing the endoscope
1. Preclean the endoscope to remove any organic debris. Follow the manufacturer’s instructions for precleaning, or
refer to established professional guidelines. See “Endoscope Precleaning and Testing” in this chapter.
2. Remove all channel valves from the endoscope and connect the ports with appropriate hookup. Refer to the
appropriate MEDIVATORS Hookup Guide for specific endoscope installation.
3. Position the endoscope in the reprocessor basin.
- Position the control section of the endoscope in the right rear of the basin.
- Position the light guide in the left front of the basin.
4. The distal end must not point upwards toward the floating lid.
- The endoscope must be completely submerged when the basin is filled.
- The endoscope must not contact the basin lid.
5. Attach the endoscope hookup connection to the basin connection. Verify there are no kinks in the hookup.
Warning! Periodically test the hookups to ensure there are no blockages and that the
connections are secure. Verify condition of Hookups and replace if worn or damaged.
Caution!
Use only MEDIVATORS supplied hookups with the DSD EDGE
®
endoscope
reprocessor.
Caution!
The hookups are not autoclavable and must be reprocessed by low temperature
disinfection only.
!
!
Summary of Contents for MEDIVATORS DSD EDGE
Page 1: ...1 Service Manual System Version 5 65...
Page 23: ...23 Figure 5 Function Keys...
Page 27: ...27 Figure 6 Set Date Screen...
Page 29: ...29 Figure 7 Set Time Screens...
Page 33: ...33 Figure 9 Display Log Screen and sample entry...
Page 42: ...42 Figure 14 Custom Program Setup Screens P1 Detergent Inject Set Minutes 00 00...
Page 43: ...43 Figure 15 Custom Program Setup Screen...
Page 51: ...51 Figure 22 Set Delayed Startup Screen...
Page 54: ...54...
Page 96: ...96 Notes...
Page 101: ...101 Figure 4 Depress locking ring Figure 5 Insert tube Figure 6 Locking ring out...
Page 115: ...115 Figure 28 Filter locations Figure 29 Disconnect the filter Figure 30 Verify inlet side...
Page 119: ...119 Figure 33 Fuse locations...
Page 121: ...121 Figure 34 Connect pressure gauge Figure 35 Bleed some air...
Page 141: ...141...
Page 142: ...142...
Page 143: ...143...
Page 149: ...149...
Page 150: ...150...
Page 151: ...151...